ICU Medical Issues a Voluntary Nationwide Recall of Certain Lots of ChemoLock(TM) and ChemoClave® Vial Spikes Due to the Potential for Burr Particulate

SAN CLEMENTE, Calif., Feb. 25, 2019 /PRNewswire/ -- ICU Medical, Inc. today announced a voluntary recall of certain lots of ChemoLock(TM) and ChemoClave(®) Vial Spikes due to the potential for burr particulate originating from the protective cap used in the assembly of the device. The affected product lot was distributed nationwide to ICU Medical customers between August 2018 and February 2019. The affected lots are listed below:


                  Item Number 
            
              Description           
           
              Lot Number

            ---

         
         CH3943              Oncology Kit w/12" Ext Set w/
                                    Spiros(TM) w/Red Cap, Clamp,
                                  Graduated Adapter; ChemoClave(TM)
                                       Vented Vial Spike, 20mm;
                                  ChemoClave(TM) Vented Vial Spike,
                                     13mm; Spiros(TM) w/Red Cap       
        3846114, 3855564, 3902719, 3961122

            ---

         
         CH3944              Oncology Kit w/12" Ext Set w/
                                    Spiros(TM) w/Red Cap, Clamp,
                                  Graduated Adapter; ChemoClave(TM)
                                       Vented Vial Spike, 20mm;
                                  ChemoClave(TM) Vented Vial Spike,
                                                 13mm                               
            3846116, 3855565

            ---

         
         CH-70S          ChemoClave(R) Vented Vial Spike, 20mm  
        3866759, 3872252, 3910046, 3910627

            ---

         
         CH-80S              ChemoClave(R) Vial Spike, 20mm     
        3826647, 3826648, 3826844, 3827598

                                                                      
        3839689, 3839690, 3854000, 3854001

                                                                      
        3866761, 3873006, 3936065, 3952512

    ---

         
         CL-80S              ChemoLock(TM) Vial Spike, 20mm    
        3757712, 3824204, 3824209, 3826652

                                                                      
        3828646, 3829557, 3836196, 3836811

                                                                      
        3853995, 3853997, 3866764, 3866766

                                                                      
        3872065, 3872072, 3872085, 3872093

                                                                      
        3872433, 3872444, 3910642, 3920773

                                                                           
            3932270, 3945868, 3946776

    ---

        
         CL-80S-10          ChemoLock(TM) Vial Spike, 20mm, 10
                                                units                 
        3820519, 3820522, 3839240, 3839241

                                                                      
        3844783, 3844784, 3854004, 3854005

                                                                      
        3910656, 3932907, 3946771, 3946772

    ---

        
         CL-80S-4           ChemoLock(TM) Closed Vial Spike w/
                                            Skirt, 4 Units                 
            3820520, 3820521, 3844785

            ---

        
         CL-80S-5           ChemoLock(TM) Closed Vial Spike w/
                                            Skirt, 5 Units            
        3820516, 3820523, 3839238, 3839239

                                                                      
        3844781, 3844782, 3853998, 3853999

                                                                      
        3866769, 3866770, 3872090, 3910645

                                                                      
        3932908, 3932910, 3946770, 3951236

    ---

          
         Z7148          ChemoClave(R) Vented Vial Spike, 20mm                
            3872008, 3899761

            ---

The FDA is classifying this as a Class 1 recall, as burr particulate has the potential to detach and, in rare circumstances, enter the fluid path causing an embolism that may damage organs or lead to a stroke, resulting in serious injury or death. To date, ICU Medical has not received any reports of adverse events related to this matter and is issuing this notification out of an abundance of caution.

ICU Medical is asking customers to discontinue the use and distribution of the affected product immediately. Check your inventory and quarantine all affected product at your facility and return the affected product to Stericycle for credit. For instructions on returning product or additional assistance, call Stericycle at 1-888-871-7114 Monday through Friday, between the hours of 8 a.m. to 5 p.m. Eastern Time.

Customers with questions regarding this recall can call ICU Medical at 1-877-946-7747, Monday through Friday, between the hours of 8 a.m. and 5 p.m. Central Time.

Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax.

    --  Complete and submit the report Online: www.fda.gov/medwatch/report.htm
    --  Regular Mail or Fax: Download form www.fda.gov/MedWatch/getforms.htm or
        call 1-800-332-1088 to request a reporting form, then complete and
        return to the address on the pre-addressed form, or submit by fax to
        1-800-FDA-0178

This recall is being conducted with the knowledge of the U.S. Food and Drug Administration.

Additional information regarding this recall, along with notifications to the customer can be found on ICU Medical's Website here.